Type
Sandwich ELISA, HRP-labelled antibody
Applications
Serum, Plasma
Sample Requirements
5 μl/well
Shipping
At ambient temperature. Upon receipt, store the product at the temperature recommended below.
Storage/Expiration
Store the complete kit at 2–8°C. Under these conditions, the kit is stable until the expiration date (see label on the box).
Calibration Curve
Calibration Range
1.56–100 U/ml
Limit of Detection
0.71 U/ml
Intra-assay (Within-Run)
n = 8; CV = 4.8%
Inter-assay (Run-to-Run)
n = 5; CV = 5.3%
Spiking Recovery
102,00%
Dilution Linearity
91,00%
Crossreactivity
- bovine Non-detectable
- cat Non-detectable
- dog Non-detectable
- goat Non-detectable
- hamster Non-detectable
- horse Non-detectable
- monkey Non-detectable
- mouse Non-detectable
- pig Non-detectable
- rabbit Non-detectable
- rat Non-detectable
- sheep Non-detectable
- chicken Not tested
- human Yes
Note
The kits are CE-IVD certified and intended for professional use.
Features
- European Union: for in vitro diagnostic use
- Rest of the world: for research use only!
- The total assay time is less than 3 hours.
- The kit measures IgG anti-IgA antibodies in serum and plasma (EDTA, citrate, heparin).
- Assay format is 96 wells.
- Calibrator is serum based.
- Quality Controls are human serum based. No animal sera are used.
- Components of the kit are provided ready to use, concentrated or lyophilized.
Research topic
Autoimmunity, Immune Response, Infection and Inflammation, Immunology, Transplantation
Summary
Anti-IgA in the IgG class are specific IgG antibodies against immunoglobulin A. Patients with IgA deficiency often demonstrate circulating antibodies against IgA, which have been suggested to be associated with transfusion reactions. Anti-IgA antibodies occur in about 10% of patients with selective IgA deficiency (IgAD) and are detected approximately in the same percentage of patients with Common Variable Immunodeficiency (CVID). Patients with anti-IgA antibodies are at risk to develop an anaphylactoid reaction to substitution gamma globulin treatment, especially upon intravenous administration.
Instructions for Use (RUO)
Instructions for Use (RUO)
Safety Information (RUO)
MSDS (RUO)
MSDS (RUO)
MSDS (RUO)
MSDS (RUO)
MSDS (RUO)
Product Brochure
CE IVD Assays
Other Documents
Declaration of Conformity
Find documents for the lot